Skip to content

Firmware Issue connecting to V3 Miniscope #15

@spencerbyers15

Description

@spencerbyers15

Hello,

Are the old versions of the DAQ firmware archived and available for download somewhere? I'm having issues recording from UCLA V3 sensors after flashing some new V3.2 DAQs with the newest firmware.

Some background on our problem-

My lab has a couple of V3.2 DAQs that were flashed with old firmware long ago, that work well with out UCLA V3 sensors. We've been using either Bonsai (with UCLAV3 nodes) or the old UCLA V3 software to record data from our miniscopes with no problem. Recently, we ordered a couple of extra DAQs to build new recording rigs. After flashing these new V3.2 DAQs with the latest firmware update (128K_EEPROM.img pre-built file, located here- https://github.com/Aharoni-Lab/Miniscope-DAQ-Cypress-firmware/tree/master/Built_Firmware) I am running into these issues recording from our V3 sensors.

  1. I can connect to the miniscope with the old miniscope V3 software (3 LED lights come on the DAQ, no flickering, red light on the miniscope CMOS, establish connection in the software and get an image), but I'm unable to set the FrameRate or Gain settings. This same pattern holds in Bonsai.
  2. This makes sense though, it says on the firmware update page- "This firmware will only work with the newest Miniscope Software so make sure to use the latest software release.". So I downloaded the latest miniscope software and tried to connect to our V3 sensors using that.
  3. Using the new software, I get some weird behavior. After running the correct config json file that is looking for a V3 sensor, it shows first that it connects to the miniscope, but then that there are frame retrieve errors and it keeps connecting and disconnecting. Along with this, the 3rd LED on the DAQ begins to flicker probably due to the unstable connection between the scope and the DAQ. I am, however, able to set the FrameRate and Gain using the new software.
    miniscope_connection_issue

Things I've tried to solve this issue-

  1. I am using a working CoAx cables that work with our DAQs flashed with old firmware.
  2. I am using a working V3 CMOS that works with our DAQs flashed with old firmware.
  3. Checked the L13 inductor on the new DAQ to see if it's burnt out. It wasn't burnt out, reads a 1.4 Ohm resistance across it like it does on a working DAQ
  4. Visual inspection of the DAQ, it looks brand new with no clear issues.

Are there any solutions to our issues other than just finding the old firmware version that works with the V3 software only? I've looked through the google groups discussion board but haven't come across this issue.

Thanks in advance for any help you can offer!