Skip to content

Commit 500f990

Browse files
Updated DAE Trouble Shooting (markdown)
1 parent b2a7edb commit 500f990

File tree

1 file changed

+1
-0
lines changed

1 file changed

+1
-0
lines changed

DAE-Trouble-Shooting.md

Lines changed: 1 addition & 0 deletions
Original file line numberDiff line numberDiff line change
@@ -470,6 +470,7 @@ The `selog.sq3` file in `C:/Data` is used by ISISICP to store log data prior to
470470
If you have multiple DAE2 VME crates then you will legitimately have multiple VXI devices in NI MAX. However there are cases when you can get two devices showing but have only one. This is usually because of either:
471471
- DAE USB cable has been plugged into a different socket in the PC
472472
- the DAE USB card has been changed
473+
473474
in these cases when the "scan for new hardware" happens on the PC, it thinks it has found a new device. usually the old device will show as VXI0 and disconnected, the enw device as VXI1. To remedy teh situation
474475
- stop visa server in NI max
475476
- delete VXI0

0 commit comments

Comments
 (0)