user-guide/50-Troubleshooting/Stirring troubleshooting.md

Lines changed: 6 additions & 0 deletions
Original file line numberDiff line numberDiff line change
@@ -49,6 +49,12 @@ Detach the top faceplate. Under the faceplate, slightly unscrew each of the four
4949
2. Try moving the fan closer to the vial holder by _gently_ tightening the screws on the bottom of the top faceplate.
5050
2. The liquid in the vial may be too viscous for the mixing to work.
5151

52+
### Stir bar goes out of balance during OD dodging (RPM changes)
53+
54+
If the stir bar stops, vibrates, or starts orbiting the vial when OD dodging changes the target RPM:
55+
1. Make sure a stirring calibration is active. Re-run it if you recently updated software or changed power/hardware. See [stirring calibration](/user-guide/hardware-calibrations#stirring-calibration-optional).
56+
2. Check that the vial is fully seated on the heating pad. A small gap (for example, from an o-ring or debris) can make the bar lose sync during RPM transitions.
57+
5258
### Stirring is not able to keep a steady RPM
5359

5460
1. Consider decreasing the value of `Kp` in the section `stirring.pid` in the config.ini by 50%.

user-guide/99-common-questions.mdx

Lines changed: 5 additions & 1 deletion
Original file line numberDiff line numberDiff line change
@@ -131,6 +131,10 @@ Stirring is performed with a micro stir bar inside the vial, controlled by a pai
131131

132132
Of course! The maximum length of a stir bar is 20mm. Keep the height low as to not interfere with the optics. (The provided stir bar is 3mm high). Try using other stir bar shapes / designs!
133133

134+
### Why does the stir bar stop or wobble when OD dodging changes RPM?
135+
136+
This is usually a calibration or seating issue. Make sure a stirring calibration is active (re-run it after software updates or hardware changes) and confirm the vial is fully seated on the heating pad. Even a small gap, for example from an o-ring or debris, can make the bar lose sync during RPM transitions.
137+
134138
### The magnets rub against the plastic screws above them, causing the stirring performance to be degraded / stop. How do I fix this?
135139

136140
Detach the top faceplate. Under the faceplate, slightly unscrew each of the four metal screws. This will add more space between the magnets and the plastic screws.
@@ -279,4 +283,4 @@ See steps [here](/user-guide/updating-software#method-2-update-using-a-zip-file-
279283

280284
### If I have multiple Pioreactor leaders on my network, what leader does `http://pioreactor.local` point to?
281285

282-
If there are multiple leaders on the same network, then each will override the other's `pioreactor.local` alias, and the alias will be unreliable. You can either access the leaders by their hostnames: `http://<leader hostname>.local`, or you can change the alias in the config.ini, too, by editing `domain_alias` under section `[ui]` (changing this requires a restart of the machine).
286+
If there are multiple leaders on the same network, then each will override the other's `pioreactor.local` alias, and the alias will be unreliable. You can either access the leaders by their hostnames: `http://<leader hostname>.local`, or you can change the alias in the config.ini, too, by editing `domain_alias` under section `[ui]` (changing this requires a restart of the machine).