Hello,
I received Calibration files for some neuropixel probes, but the quality of the signal I am getting is clearly worse than it should be, and I wonder if it would be possible to help me understand where these issues are coming from. Every other channel is severely clipped in the LFP, apparently owing to a too-high gain (see image of LFP band and AP band below). In fact, this is what the signal looked like before it had a calibration file.
For some reason, one half of the probe is very noisy, and the other half is not as noisy, though it is getting somewhat clipped, which is in turn impacting the AP band. I wonder if this is a problem with the calibration file. I have attached the Calibration file, and below are some screenshots of the LFP and AP signal from the Neuropixel PXI tab in the open ephys acquisition software. 18005123662.zip
This signal is from a mouse moving freely in an open arena, sometimes bumping its cap into the side of the wall. We used the Buzsaki microdrive and cap design found here, the ground and reference wires are shorted, and I have a chlorided ground wire in the brain. I would be very grateful for any feedback you might have!