Sysmex XN-L hematology analyzer ASTM interface plugin.
- ASTM Pattern: H-segment field 4 contains
XN-L
| Property | Value |
|---|---|
| Protocol | ASTM LIS2-A2 or ASTM E1381-02 |
| Transport | RS232 Serial or TCP/IP |
| Category | HEMATOLOGY |
| Manufacturer | Sysmex |
Supports 30+ CBC parameters including extended differential and reticulocyte counts:
| Analyzer Code | OpenELIS Test | Units |
|---|---|---|
| WBC | White Blood Cell Count | 10^3/µL |
| RBC | Red Blood Cell Count | 10^6/µL |
| HGB | Hemoglobin | g/dL |
| HCT | Hematocrit | % |
| MCV | Mean Corpuscular Volume | fL |
| MCH | Mean Corpuscular Hemoglobin | pg |
| MCHC | Mean Corpuscular Hemoglobin Concentration | g/dL |
| PLT | Platelet Count | 10^3/µL |
| NEUT_COUNT, LYMPH_COUNT, MONO_COUNT, EO_COUNT, BASO_COUNT | 5-Part Differential Counts | 10^3/µL |
| NEUT_PERCENT, LYMPH_PERCENT, MONO_PERCENT, EO_PERCENT, BASO_PERCENT | 5-Part Differential Percentages | % |
| RET_COUNT, RET_PERCENT | Reticulocyte Counts | Various |
| ... | (Additional advanced parameters) | Various |
- Protocol: ASTM LIS2-A2 or ASTM E1381-02
- Framing: ENQ/ACK/NAK establishment
- Transport: RS232 (9600 baud, 8N1) or TCP/IP
cd plugins/analyzers/SysmexXN-L
mvn clean packageCopy the built JAR to /var/lib/openelis-global/plugins/ and restart OpenELIS.
Supports both RS232 and TCP/IP connectivity per vendor specifications. For file-based XN series, see SysmexXN1000.
Last Verified: 2026-02-02