Medtronic MyCareLink Patient Monitor, 24950 MyCareLink...
Moderate severity
Unreviewed
Published
May 13, 2022
to the GitHub Advisory Database
•
Updated May 22, 2025
Description
Published by the National Vulnerability Database
Jul 3, 2018
Published to the GitHub Advisory Database
May 13, 2022
Last updated
May 22, 2025
Medtronic MyCareLink Patient Monitor, 24950 MyCareLink Monitor, all versions, and 24952 MyCareLink Monitor, all versions, contains debug code meant to test the functionality of the monitor's communication interfaces, including the interface between the monitor and implantable cardiac device. An attacker with physical access to the device can apply the other vulnerabilities within this advisory to access this debug functionality. This debug functionality provides the ability to read and write arbitrary memory values to implantable cardiac devices via inductive or short range wireless protocols. An attacker with close physical proximity to a target implantable cardiac device can use this debug functionality.
References