Skip to content

The Hospira LifeCare PCA Infusion System before 7.0 does...

High severity Unreviewed Published May 17, 2022 to the GitHub Advisory Database • Updated Apr 12, 2025

Package

No package listedSuggest a package

Affected versions

Unknown

Patched versions

Unknown

Description

The Hospira LifeCare PCA Infusion System before 7.0 does not validate network traffic associated with sending a (1) drug library, (2) software update, or (3) configuration change, which allows remote attackers to modify settings or medication data via packets on the (a) TELNET, (b) HTTP, (c) HTTPS, or (d) UPNP port. NOTE: this issue might overlap CVE-2015-3459.

References

Published by the National Vulnerability Database Jul 6, 2015
Published to the GitHub Advisory Database May 17, 2022
Last updated Apr 12, 2025

Severity

High

EPSS score

Exploit Prediction Scoring System (EPSS)

This score estimates the probability of this vulnerability being exploited within the next 30 days. Data provided by FIRST.
(69th percentile)

Weaknesses

Insufficient Verification of Data Authenticity

The product does not sufficiently verify the origin or authenticity of data, in a way that causes it to accept invalid data. Learn more on MITRE.

CVE ID

CVE-2014-5406

GHSA ID

GHSA-vxf4-qh89-w2r6

Source code

No known source code

Dependabot alerts are not supported on this advisory because it does not have a package from a supported ecosystem with an affected and fixed version.